Fast stopped-flow microcalorimeter

Journal of Biochemical and Biophysical Methods(1981)

引用 6|浏览7
暂无评分
摘要
A fast stopped-flow microcalorimeter using a thermistor and an a.c. bridge is described. Time resolution, limited by the response of the thermistor, is 3 ms for t12 while sensitivity is 100 μ°C in this bandpass. A reaction requires 0.3 ml of each reactant. The microcalorimeter is adiabatic to within 2% for 2 s and mixing may be repeated every 5 min. Computer finite element simulation techniques (FEST) are used to correct for the thermistor response time and heat losses. Results of tests with reactions of NaOH and HCl, NaHCO3 and HCl, and glycylglycine and CO2 agree with published kinetic and thermodynamic values to within an experimental error of less than 2%.
更多
查看译文
关键词
Stopped-flow microcalorimetry,fast thermistors,carbon dioxie,glycylglycine,finite element simulation