Efficiency of Pt and Ir modifiers for the simultaneous determination of As, Se and In in a sodium sulphate matrix by electrothermal atomic absorption spectrometry

Talanta(2003)

引用 9|浏览1
暂无评分
摘要
The effect of pre-reduced Pt and Ir modifiers towards simultaneous determination of As, Se and In in a sodium sulphate matrix was investigated. In spite of application of very isothermal THGA atomizer, negative influence of the matrix is pronounced. The efficiency of iridium modifier is somewhat higher than that of platinum modifier. However compared to palladium modifier both are significantly less effective in sulphate media. Taking into account our previous data, effectiveness of platinum group metal (PGM) modifiers in the determination of As, Se and In in the presence of sodium sulphate increases in the order Ru更多
查看译文
关键词
Electrothermal atomic absorption spectrometry,Chemical modifiers,Sodium sulphate matrix,Platinum,Iridium,Selenium,Arsenic,Indium