Impact On Hearing Aid Targets Of Measuring Thresholds In Db Hl Versus Db Spl

INTERNATIONAL JOURNAL OF AUDIOLOGY(2003)

引用 20|浏览1
暂无评分
摘要
Audiometric measurements are traditionally made in dB HE, which by definition are specified relative to the sound pressure level (SPL) in a coupler. Real-ear dB SPL is then estimated by applying an average ear transform to the coupler value. However, individual variation in ear canal acoustics and variations in transducer placement strongly influence the dB SPL of signals arriving at the eardrum. In this paper, data from 1814 ears are presented, showing that the distribution of eardrum dB SPL for a fixed signal level varies across ears and across frequency by as much as 40 dB. The impact of this variance upon hearing aid targets computed with the NAL-NL1 fitting algorithm is examined by comparing the targets obtained from using an average transform with those obtained when audiometric data in dB SPL are obtained by applying individually measured real-ear-to-coupler difference (RECD) values to dB HL thresholds. The impact can be considerable.
更多
查看译文
关键词
hearing aid fittings, dB SPL versus dB HL, real-ear measurement