A dynamic model for the estimation of optimum timing of computed tomography scan for dose evaluation of 125I or 103Pd seed implant of prostate

International Journal of Radiation Oncology*Biology*Physics(1999)

引用 59|浏览8
暂无评分
摘要
Purpose/Objective: The dosimetric evaluation of permanent 125I or 103Pd prostate implant is based on the assumption that both prostate and seeds are static throughout the entire treatment time which lasts months. However, the prostate is often edematous after the surgical implantation of seeds. Therefore, both the volume of the prostate and the seed locations change dynamically as the edema resolves. This effect has impact on the validity of postimplant analysis based upon a CT scan. If a CT scan is taken too early after implantation while there is edema in the prostate, the dose delivered by the implant may be underestimated. If the imaging is delayed too long, the dose may be overestimated. The magnitude of this effect depends on both of the half-life of the isotope used and the half-life and magnitude of the edema. This study describes a dynamic biomathematical model which takes edema into account in calculating the dose delivered by the implant and is used to investigate the optimum time to obtain the postimplant CT scan.
更多
查看译文
关键词
Brachytherapy,Prostate implants,125I,103Pd,Edema,Post-implant dosimetry