Evaluation of transmission methodology and attenuation correction for the microPET Focus 220 animal scanner.

PHYSICS IN MEDICINE AND BIOLOGY(2006)

引用 34|浏览10
暂无评分
摘要
An accurate, low noise estimate of photon attenuation in the subject is required for quantitative microPET studies of molecular tracer distributions in vivo. In this work, several transmission-based measurement techniques were compared, including coincidence mode with and without rod windowing, singles mode with two different energy sources (Ge-68 and Co-57), and postinjection transmission scanning. In addition, the effectiveness of transmission segmentation and the propagation of transmission bias and noise into the emission images were examined. The Co-57 singles measurements provided the most accurate attenuation coefficients and superior signal-to-noise ratio, while Ge-68 singles measurements were degraded due to scattering from the object. Scatter correction of Ge-68 transmission data improved the accuracy for a 10 cm phantom but over-corrected for a mouse phantom. Co-57 scanning also resulted in low bias and noise in postinjection transmission scans for emission activities up to 20 MBq. Segmentation worked most reliably for transmission data acquired with Co-57 but the minor improvement in accuracy of attenuation coefficients and signal-to-noise may not justify its use, particularly for small subjects. We conclude that Co-57 singles transmission scanning is the most suitable method for measured attenuation correction on the microPET Focus 220 animal scanner.