Sagittal and vertical load-deflection and permanent deformation of transpalatal arches connected with palatal implants: An in-vitro study

American Journal of Orthodontics and Dentofacial Orthopedics(2007)

引用 8|浏览6
暂无评分
摘要
Introduction: The purposes of this laboratory investigation were to (1) measure the sagittal and vertical deflection of loaded transpalatal arches (TPAs) connected to a palatal implant, (2) measure the extent of permanent deformation of the connecting TPA in the sagittal and vertical directions, (3) test various wire dimensions in terms of deflection behavior, and (4) evaluate soldering vs laser welding vs adhesive bonding of TPAs in terms of load deflection behavior. Methods: Stainless steel wires of 6 dimensions were tested: 0.8 X 0.8, 0.9, 1, 1.1, 1.2, and 1.2 X 1.2 mm. For each dimension, 10 specimens were soldered to the palatal implant abutment, 10 were laser welded, and 10 were adhesively bonded to the implant abutment (total, 180 specimens). The measuring device applied increments of force of 50 cN, from 0 to 500 cN. Then the specimens were unloaded. The values were statistically described and analyzed with ANOVA and Wilcoxon rank sum tests. Results and Conclusions: Absolute orthodontic anchorage without deformation of TPAs was not observed with the wire dimensions tested. To prevent loss of anchorage greater than 370 mu m (sagittal deflection of 1.2 X1.2 mm adhesively bonded TPA at 500 cN force level), wires thicker than 1.2 X 1.2 mm or cast anchorage elements must be considered for clinical practice. However, larger cross sections might cause more patient discomfort, and laboratory procedures increase costs.
更多
查看译文
关键词
transpalatal implants,transpalatal arches,permanent deformation,load-deflection,in-vitro
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要