ESTUDO ELETROQUÍMICO COMPARATIVO DO AÇO ISO 5832-9 EM DIFERENTES MEIOS DE INTERESSE BIOLÓGICO

REM-REVISTA ESCOLA DE MINAS(2010)

引用 0|浏览2
暂无评分
摘要
The electrochemical behavior of ISO 5832-9 stainless steel at 37 degrees C in 0.9% NaCl, Ringer Lactate and minimum essential medium (MEM) has been studied, using linear voltammetry, and surface analysis by SEM and EDS. Mechanical and toxicity tests were made. ISO 5832-9 is passivated at corrosion potential (E) and it does not present pitting corrosion on the media studied from to 50 in V above the transpassivation potential (Ei). SEM and EDS analysis have shown that the sample previously immersed in MEM presents a diffirent behavior at 50 in V above El: the manganese oxide inclusions are absent in the surface. E. values and passivation current density values j(pass) changed according to the following. E(corr, RL) < E(corr,NaCl) < E(corr, MEM) and J (MEM) << j(RL) congruent to j(NaCl) The stainless steel was characterized as non toxic in the cytotoxicity assay
更多
查看译文
关键词
ISO 5832-9 stainless steel,orthopedic implant,corrosion,physiological medium