Separation And Determination Of Organogermanium Compounds By Ion Chromatography
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES(1997)
摘要
For nearly three decades, organogermanium compounds have become increasingly of interest owing to their extensive physiological and pharmaceutical activity. In this paper, two new high performance ion chromatographic methods for separation and determination of three kinds of organogermanium compounds beta-carboxyethylgermanium sesquioxide (I), beta-(alpha-methyl)-carboxyethylgermanium sesquioxide (II) and d-(beta-carboxyethyl)germanium hydroxide (III) were proposed. A Dionex DX-300 ion chromatograph equipped with a Dionex FED-II pulsed electrochemical detector (conductivity mode), and a Dionex AI-450 chromatography workstation was employed. The separation was achieved by using ion-exchange or ion-exclusion mechanism.The detection limits(S/N=3, expressed as germanium) for the three compounds were all below sub- mu g/mL level. The methods have been applied to the analysis of tonic oral drinks, and the average recoveries for the three compounds range from 95 - 108%. The results obtained were in agreement with those of hydride generation atomic fluorescence spectrometry (HG-AFS).
更多查看译文
关键词
detection limit,ion chromatography,ion exchange