Contamination from electrically conductive silicone tubing during aerosol chemical analysis

Atmospheric Environment(2009)

引用 20|浏览6
暂无评分
摘要
Electrically conductive silicone tubing is used to minimize losses in sampling lines during the analysis of airborne particle size distributions and number concentrations. We report contamination from this tubing using gas chromatography–mass spectrometry (GC–MS) of filter-collected samples as well as by particle mass spectrometry. Comparison of electrically conductive silicone and stainless steel tubing showed elevated siloxanes only for the silicone tubing. The extent of contamination increased with length of tubing to which the sample was exposed, and decreased with increasing relative humidity.
更多
查看译文
关键词
Aerosol chemical analysis,Electrically conductive silicone tubing,Siloxanes,Contamination