Mass Accuracy and Sequence Requirements for Protein Database Searching

Analytical Biochemistry(1999)

引用 31|浏览10
暂无评分
摘要
To elucidate the role of high mass accuracy in mass spectrometric peptide mapping and database searching, selected proteins were subjected to tryptic digestion and the resulting mixtures were analyzed by electrospray ionization on a 7 Tesla Fourier transform mass spectrometer with a mass accuracy of 1 ppm. Two extreme cases were examined in detail: equine apomyoglobin, which digested easily and gave very few spurious masses, and bovine α-lactalbumin, which under the conditions used, gave many spurious masses. The effectiveness of accurate mass measurements in minimizing false protein matches was examined by varying the mass error allowed in the search over a wide range (2–500 ppm). For the “clean” data obtained from apomyoglobin, very few masses were needed to return valid protein matches, and the mass error allowed in the search had little effect up to 500 ppm. However, in the case of α-lactalbumin more mass values were needed, and low mass errors increased the search specificity. Mass errors below 30 ppm were particularly useful in eliminating false protein matches when few mass values were used in the search. Collision-induced dissociation of an unassigned peak in the α-lactalbumin digest provided sufficient data to unambiguously identify the peak as a fragment from α-lactalbumin and eliminate a large number of spurious proteins found in the peptide mass search. The results show that even with a relatively high mass error (0.8 Da for mass differences between singly charged product ions), collision-induced dissociation can help identify proteins in cases where unfavorable digest conditions or modifications render digest peaks unidentifiable by a simple mass mapping search.
更多
查看译文
关键词
electrospray ionization,collision induced dissociation,mass spectrometer,database search
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要