Addition of a Defibrillation Electrode in the Low Right Atrium to a Right Ventricular Lead Does Not Reduce Ventricular Defibrillation Thresholds.

Pacing and clinical electrophysiology(2004)

引用 8|浏览7
暂无评分
摘要
Transvenous unipolar active can defibrillation systems have proven to be effective in treating ventricular tachyarrhythmias. However, a further reduction of ventricular defibrillation thresholds (V‐DFT) would increase the longevity, reduce the size of pulse generators, and help to avoid additional leads in patients with inacceptable high V‐DFTs. In a finite difference computer model, the extension of the right ventricular (RV) defibrillation coil into the low right atrium led to a 40% reduction of unipolar V‐DFT. To evaluate this finding, we conducted a prospective, randomized study in 11 patients receiving an ICD. Extension of the RV electrode was simulated by adding a second coil placed in the low right atrium with the same polarity. Using a binary search protocol, V‐DFT was determined with and without the additional electrode in each patient. Total shock impedance was significantly lower in the two coil (low RA) configuration, compared to the single coil (RV) configuration. Corresponding values were 49.9 ± 6.7 Ohm and 61.1 ± 9.3 Ohm, respectively (P < 0.01, paired t‐test). However, there was no reduction, but even a nonsignificant increase in V‐DFTs. Mean V‐DFT in the RV configuration was 12.0 ± 5.6 J and 16.3 ± 7.8 J in the low RA configuration (P = 0.09, paired t‐test). Despite a reduction in total impedance, the addition of a defibrillation coil in the low right atrium does not reduce ventricular defibrillation thresholds. (PACE 2004; 27:346–351)
更多
查看译文
关键词
implantable cardioverter defibrillator,defibrillation threshold,leads,lead position
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要