Titanium(IV)/cryptand22 and Zinc(II)/cryptand22-coated piezoelectric (PZ) quartz crystal detection systems with home-made computer interfaces for data acquisition and processing were developed and successfully applied for the detection of nitrogen dioxide (NO2) and carbon monoxide (CO), respectively. Among the 20 metal ion/cryptand22 complexes, e.g. Mn2+/cryptand22, Ti4+/cryptand22, Zn2+/cryptand22, Al3+/cryptand22, Cr3+/cryptand22, Cu2+/cryptand22 and Ru3+/cryptand22, Ti4+/cryptand22 and Zn2+/cryptand22 were found to be the best adsorbents to adsorb the NO2 and CO in the air respectively. The Ti4+/cryptand22 coated PZ crystal NO2 sensor exhibited a good linear response to the concentration of NO2 with excellent sensitivity (9.10Hz/ppm), detection limit (0.16ppm NO2) and reproducibility with relative standard deviation (RSD) of 2.09%. The Zn2+/cryptand22 coated CO sensor also showed a good sensitivity of approximately 1.6Hz/ppm to CO with a detection limit of 0.91ppm and reproducibility with RSD of 2.15%. The coating load effects on the frequency responses of these cryptand22/metal ion coated PZ crystal NO2 and CO sensors were also investigated. The frequency responses of the PZ crystal sensors increased with the coating load. The cryptand22/metal ion coated PZ crystal sensors were also successfully applied as gas chromatographic (GC) detectors to detect NO2, CO and some organic compounds, e.g. propyl amine and propanoic acid. The Ti4+/cryptand22 coated PZ crystal GC detector compared well with the thermal conductivity detector (TCD) for NO2 vapor which TCD did not response. An artificial back propagation neural (BPN) network with a two-channel PZ crystal flow sensor was prepared and applied to recognize and distinguish each component (NO2 and CO) in a mixed NO2/CO sample and it showed quite good distinction of NO2 and CO in the mixed gas sample. The concentrations of NO2 and CO were also computed by multivariate linear regression analysis (MLR).
更多
查看译文
关键词
NO2 sensor,CO sensor,Piezoelectric crystal,Cryptand,Back propagation neural (BPN) network,Multivariate linear regression (MLR) analysis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要