Simultaneous measurement of cellular respiration and acidification with a single CMOS ISFET.

Biosensors and Bioelectronics(2001)

引用 130|浏览17
暂无评分
摘要
In vivo, the pH value and oxygen partial pressure are the most important physico–chemical parameters in the microenvironment of human tissues. In vitro, the extracellular acidification rate of cell cultures is an indicator of global cellular metabolism, while the rate of oxygen consumption is a measure of mitochondrial activity. Earlier approaches had the disadvantage that these two values had to be measured with two separate sensors at different loci within the tissue or cell culture. Furthermore, conventional Clark-type oxygen sensors are not very compatible for miniaturisation, making it impossible to measure at small cell volumes or even at the single cell level. We have, therefore, developed an ISFET based sensor structure which is able to measure both pH and oxygen partial pressure. This sensor structure was tested in vitro for simultaneous records of cellular acidification and respiration rates at the same site within the cell culture. This sensor is manufactured by a CMOS-process.
更多
查看译文
关键词
Acidification,Cellular sensor,Heterogeneity,ISFET,Respiration