Assessment of radiation exposure outside the radiotherapeutic room during medical accelerator beam emission with the use of TL detectors (radiation exposure outside a LINAC room).

RADIATION PROTECTION DOSIMETRY(2013)

引用 2|浏览5
暂无评分
摘要
Photon and neutron soft tissue absorbed doses near the entrance door to the medical LINAC treatment room were measured with the use of thermoluminescent detectors LiF:Mg,Cu,P in the anthropomorphic phantom. Two different therapeutic beams (6 and 15 MV) and four treatment techniques were involved in the present study. This allowed one to investigate the contribution of scattered X rays, secondary neutrons and gamma rays to the radiation field. Photon absorbed dose rates 50 cm away from the LINAC room door during emission of 15-MV X rays varied between 4.1x10(-4) and 5.6x10(-4) Gy h(-1), depending on the gantry position and the irradiation field size, whereas in the case of 6-MV therapeutic irradiation these doses are similar to 1.5 times lower. In the case of 15-MV beam emission, a mixed radiation field near the bunker door is observed with the photon radiation as the main component, which includes a 33.1 % contribution of the induced gamma radioactivity and similar to 2.1 % contribution of the neutron radiation.