Assessment of radiation exposure outside the radiotherapeutic room during medical accelerator beam emission with the use of TL detectors (radiation exposure outside a LINAC room).
RADIATION PROTECTION DOSIMETRY(2013)
摘要
Photon and neutron soft tissue absorbed doses near the entrance door to the medical LINAC treatment room were measured with the use of thermoluminescent detectors LiF:Mg,Cu,P in the anthropomorphic phantom. Two different therapeutic beams (6 and 15 MV) and four treatment techniques were involved in the present study. This allowed one to investigate the contribution of scattered X rays, secondary neutrons and gamma rays to the radiation field. Photon absorbed dose rates 50 cm away from the LINAC room door during emission of 15-MV X rays varied between 4.1x10(-4) and 5.6x10(-4) Gy h(-1), depending on the gantry position and the irradiation field size, whereas in the case of 6-MV therapeutic irradiation these doses are similar to 1.5 times lower. In the case of 15-MV beam emission, a mixed radiation field near the bunker door is observed with the photon radiation as the main component, which includes a 33.1 % contribution of the induced gamma radioactivity and similar to 2.1 % contribution of the neutron radiation.