Biomechanical effects of plate area and locking screw on medial open tibial osteotomy.

COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING(2015)

引用 23|浏览12
暂无评分
摘要
Medial open high tibial osteotomy (HTO) has been used to treat osteoarthritis of the medial compartment of the knee. However, weaker plate strength, unstable plate/screw junction and improper surgery technique are highly related to the HTO outcomes. Two pi-shape plates were designed and eight variations (two supporting area x four locking stiffness) were compared by finite-element method. The computed tomography-based tibia was reconstructed and both wedge micromotion and implant stresses were chosen as the comparison indices. The construct was subjected to surgical and physiological loads. The medial-posterior region is the most loaded region and the load through the posterior leg is about four times that through the anterior leg. This indicates that the two-leg design can form a force-couple mechanism to effectively reduce the implant stresses. The use of locking screws significantly decrease the screw and hole stresses. However, the extending plate reduces the stresses of screws and holes above the wedge but makes the distal screws and holes much stressed. Wedge micromotion is affected by extending plate rather than locking screw. Three factors contribute to effective stabilisation of unstable HTO wedge: (1) intimate tibia-plate contact at medial-posterior regions, (2) sufficient rigidity at plate-screw junctions and (3) effective moment-balancing design at distal tibia-plate interfaces.
更多
查看译文
关键词
finite-element analysis,high tibial osteotomy,plate,locking screw,knee