A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads

SCIENTIFIC REPORTS(2015)

引用 9|浏览69
暂无评分
摘要
Active-fixation pacing leads allow the use of selective pacing sites. We evaluated their long-term performance versus passive-fixation leads in 199 newly implanted patients (n = 100 active and n = 99 passive). Postoperative pacing thresholds in the active group were higher than in the passive group (0.85 ± 0.31 V vs. 0.53 ± 0.21 V at baseline, P < 0.001). The active thresholds fell to 0.72 ± 0.23 V at 5 years with a significant drop at one month (0.68 ± 0.53 V, P = 0.003). The passive thresholds slightly increased to 0.72 ± 0.31 V at five years. Differences between groups were significant until three years (all P < 0.05). Active impedances were generally lower than passive impedances (600.44 ± 94.31Ω vs. 683.14 ± 110.98Ω at baseline) and both showed significant reductions at one month to 537.96 ± 147.43Ω in the active group and after three months to 643.85 ± 82.40Ω in the passive group (both P < 0.01 vs. baseline). Impedance differences between groups were significant until four years (all P < 0.05). Adverse events included thresholds over 1 V, 5 of 6 active and 2 of 5 passive leads returned to below 1 V. One active left ventricular lead dislodged. One passive left subclavian lead insulation fracture occurred. Thus Active fixation pacing leads are stable in a five-year long-term follow up. There was no difference between active and passive leads in terms of electrical performance.
更多
查看译文
关键词
Cardiovascular diseases,Medical research,Science,Humanities and Social Sciences,multidisciplinary
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要