Reduction of fetal dose in computed tomography using anterior shields.

JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY(2015)

引用 8|浏览1
暂无评分
摘要
Objective: The objective of this study was to assess the potential of anterior in beam bismuth shields placed on the abdomen and the pelvis of a RANDO phantom to decrease radiation dose to a simulated fetus. Methods: Thermal luminescent dosimeters irradiated in a RANDO phantom were used to measure radiation dose at 0 weeks and 18 weeks after conception. An 18-week gestation was simulated by adding the corresponding equivalent of super-stuff to account for additional subcutaneous tissue. To assess image quality, a 461A body phantom with low contrast resolution phantom insert was used. The minimum size of low contrast resolution target visualize was recorded with no shielding and 25-cm-wide shielding. Results: With shields, there was a radiation dose reduction of 19% to 23% at 0-week gestation and 16% to 24% at 18-week gestation. No qualitative difference was found in low contrast detectability with the shields. Conclusions: These shields reduce fetal radiation exposure early in pregnancy with acceptable image quality.
更多
查看译文
关键词
bismuth shielding,fetal radiation dose