Effective Management Of Atrioventricular Interval For Paroxysmal Atrial Fibrillation That Developed After Dddr Pacemaker Implantation In A Sick Sinus Syndrome Patient

INTERNATIONAL HEART JOURNAL(2015)

引用 0|浏览15
暂无评分
摘要
A 68-year-old man with sick sinus syndrome (SSS) was referred to our department for pacemaker implantation. After implantation of a pacemaker with rate-responsive dual chamber (DDDR) mode and minimized ventricular pacing (MVP) functions, paroxysmal atrial fibrillation (PAF) repeatedly developed. Pacemaker memory showed that the intrinsic atrioventricular (AV) (atrial pacing-ventricular sensing [Ap-Vs]) interval was paradoxically prolonged during rate-responsive atrial single-chamber (AAIR) mode rapid pacing because of MVP. Accordingly, to eliminate the paradoxical prolongation of the AV interval during rapid atrial pacing, we changed MVP to medium AV hysteresis and conducted DDDR mode pacing with rate-dependent AV delay. PAP then sharply decreased without antiarrhythmic drugs.
更多
查看译文
关键词
Minimized ventricular pacing, Intrinsic rhythm support plus, Medium AV hysteresis, Rate-response, Closed-loop stimulation, Intrinsic AV interval, Paradoxical AV prolongation, Rate-dependent AV delay