谷歌浏览器插件
订阅小程序
在清言上使用

Precise Determination of Bromine in Pp Resin Pellet by Instrumental Neutron Activation Analysis Using Internal Standardization

Journal of radioanalytical and nuclear chemistry(2014)

引用 7|浏览16
暂无评分
摘要
Instrumental neutron activation analysis with the internal standardization was applied to the precise determination of Br in polypropylene resin of candidate certified reference material. The known amount of 197Au was used as an internal standard to compensate for neutron flux inhomogeneity, to improve the γ ray measurement uncertainty and the linearity of the calibration curves. The reliability of the proposed method validated using analytical results of BCR-681. The analytical result of Br in the sample was consistent with that obtained by ID-ICPMS. The relative expanded uncertainty (k = 2) was 1.5 %, and it was equivalent to that of ID-ICPMS.
更多
查看译文
关键词
Neutron activation analysis,Bromine,Internal standard method,RoHS directive,Measurement uncertainty,Certified reference material