Construction and calibration of the multi-neutron correlation spectrometer at Peking University

H B You,Z H Yang,Y L Ye,Z H Li,Y C Ge,J L Lou,Q T Li, Jiening Xiao,Rui Qiao, Z Y Tian, Yan Lindsay Sun,H N Liu, Jian Chen,Junjie Wu,Bo Yang, Ke Yang,J S Wang,Y Y Yang,Pengfei Ma,J B Ma,S L Jin, J L Han

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2013)

引用 10|浏览39
暂无评分
摘要
A Multi-Neutron Correlation Spectrometer (MuNCoS) composed of 80 detector modules was constructed at Peking University. A position resolution of less than 3cm (FWHM) for a typical long scintillation module was obtained based on cosmic ray tests. A calibration method using an unstable nucleus beam of 11Be, together with the fragment identification by using a zero degree telescope, has been successfully developed. Neutron detection efficiencies of 15(1)% and 17(2)% were achieved for neutron energy intervals of 30–40MeV and 40–60MeV, respectively, and at a signal charge threshold of 4MeVee. The experimentally determined neutron detection efficiencies agree well with the simulation based on the GEANT4 code.
更多
查看译文
关键词
Neutron detection,Position resolution,Intrinsic efficiency,Unstable nucleus