Consistency in quality correction factors for ionization chamber dosimetry in scanned proton beam therapy.

MEDICAL PHYSICS(2017)

引用 12|浏览16
暂无评分
摘要
Purpose: The IAEA TRS-398 code of practice details the reference conditions for reference dosimetry of proton beams using ionization chambers and the required beam quality correction factors (kQ). Pencil beam scanning (PBS) systems cannot approximate reference conditions using a single spot. However, dose distributions requested in TRS-398 can be reproduced with PBS using a combination of spots. This study aims to demonstrate, using Monte Carlo (MC) simulations, that kQ factors computed/measured for broad beams can be used with scanned beams for similar reference dose distributions with no additional significant uncertainty. Methods: We consider the Alfonso formalism 13 usually employed for nonstandard photon beams. To approach reference conditions similar as IAEA TRS-398 and the associated dose distributions, PBS must combine many pencil beams with range or energy modulation and shaping techniques that differ from those used in passive systems (broad beams). In order to evaluate the impact of these differences on kQ factors, ionization chamber responses are computed with MC (Geant4 9.6) in three different proton beams, with their corresponding quality factors (Q), producing a 10 9 10 cm 2 field with a flat dose distribution for (a) a dedicated scanned pencil beam (Q(pbs)), (b) a hypothetical proton source (Qhyp), and (c) a double-scattering beam (Q(ds)). The tested ionization chamber cavities are a 2 x 2 x 0.2 mm 3 air cavity, a Roos-type ionization chamber, and a Farmer-type ionization chamber. Results and Discussion: Ranges of Q(pbs), Qhyp, and Qds are consistent within 0.4 mm. Flatnesses of dose distributions are better than 0.5%. Calculated k fpbs; fref Q(pbs); Q(hyp) is 0.999 + 0.002 for the air cavity and the Farmer-type ionization chamber and 1.001 + 0.002 for the Roos-type ionization chamber. The quality correction factors k fpbs; fref Q(pbs); Q(ds) is 0.999 + 0.002 for the Farmer-type and Roos-type ionization chambers and 1.001 + 0.001 for the Roos-type ionization chamber. Conclusion: The Alfonso formalism was applied to scanned proton beams. In our MC simulations, neither the difference in the beam profiles (scanned beam vs hypothetical beam) nor the different incident beam energies influenced significantly the beam correction factors. This suggests that ionization chamber quality correction factors in scanned or broad proton beams are indistinguishable within the calculation uncertainties provided dose distributions achieved by both modalities are similar and compliant with the TRS-398 reference conditions. (C) 2017 American Association of Physicists in Medicine
更多
查看译文
关键词
beam quality correction factors,pencil beam scanning,proton therapy,reference dosimetry
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要