Human erythrocyte lifespan measured by Levitt's CO breath test with newly developed automatic instrument.

JOURNAL OF BREATH RESEARCH(2018)

引用 40|浏览24
暂无评分
摘要
Existing standard techniques for erythrocyte (RBC) lifespan measurement, such as quantitation of labeling with isotopes or biotin, are cumbersome and time-consuming. Given that endogenous CO originates mainly from degraded RBCs, a team lead by Levitt developed a CO breath test to enable more efficient RBC lifespan estimation. The purpose of this study was to evaluate the reliability of Levitt's CO breath test method with our newly developed automatic instrument. RBC lifespan measurements conducted by Levitt's CO breath test method were conducted in 109 healthy subjects and 91 patients with chronic hemolytic anemia. In healthy subjects, the RBC lifespan was 126 +/- 26 days, similar to values obtained with classical standard labeling methods. RBC lifespan did not differ significantly between males and females or between juveniles and adults, and did not correlate with age. To our knowledge, this datum represents an RBC lifespan average for the largest sample to date. In subjects with hemolytic anemia, RBC lifespan was 29 +/- 14 days, which is significantly shorter than that of the healthy subjects (p = 0.001). Using 75 days as a cut-off, diagnostic accuracy for hemolytic anemia in the present study sample was 100%. In conclusion, the present results indicate that Levitt's CO breath test is an ideal method for human RBC lifespan measurement, and the newly developed automatic instrument is reliable and convenient for clinical practice.
更多
查看译文
关键词
human red blood cell lifespan,alveolar endogenous CO,non-dispersive infrared spectroscopy,hemolytic anemia
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要