Arterial inflow line equipped with a side arm for circulatory support and catheter insertion during transcatheter aortic valve implantation for limited vascular access

General thoracic and cardiovascular surgery(2018)

引用 1|浏览19
暂无评分
摘要
Transcatheter aortic valve implantation was performed on a 78-year-old patient. Elective circulatory support with cardiopulmonary bypass was planned because of left ventricular function impairment and hemodynamic instability. Limited vascular access was due to a severe atherosclerotic aorta distal to the origin of the left carotid artery. The right arm was the only safe vascular access site. However, at least 2 vascular access sites for angiographic catheter and inflow of circulatory support were required. An arterial inflow line equipped with a side arm was developed to enable single access to the right axillary artery to be used for the above purposes.
更多
查看译文
关键词
Axillary artery cannulation,Cardiopulmonary bypass,Shaggy aorta,Transcatheter aortic valve implantation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要