Preliminary Report on Stone Breakage and Lesion Size Produced by a New Extracorporeal Electrohydraulic (Sparker Array) Discharge Device.

Urology(2018)

引用 1|浏览18
暂无评分
摘要
OBJECTIVE To determine if an innovative extracorporeal electrohydraulic shock wave (SW) device (sparker array [SPA]) can effectively fracture artificial stones in vitro and in vivo, and if SPA treatment produces a renal lesion in our pig model of lithotripsy injury. Results of these experiments will be used to help evaluate the suitability of this device as a clinical lithotripter. MATERIALS AND METHODS Ultracal-30 artificial stones were placed in a holder at the focus of the SPA and treated with 600 SWs (21.6 kV, 60 shocks/min). Stone fragments were collected, dried, and weighed to determine stone breakage. In vivo stone breakage entailed implanting stones into pigs. These stones were treated with 600 or 1200 SWs and the fragments were collected for analysis. Lesion analysis consisted of treating the left kidney of pigs with 1200 or 2400 SWs and quantitating the hemorrhagic lesion. RESULTS In vitro, 71% +/- 2% of each artificial stone was fractured to < 2 mm in size. In vivo stone breakage averaged 63%. Renal injury analysis revealed that only 1 of 7 kidneys showed evidence of hemorrhagic injury in the treated area. CONCLUSION The SPA consistently comminuted artificial stones demonstrating its ability to fracture stones like other lithotripters. Also, the SPA caused little to no renal injury at the settings used in this study. These findings suggest further research is warranted to determine the potential of this device as a clinical lithotripter. (C) 2018 Elsevier Inc.
更多
查看译文
关键词
acute kidney injury,artificial stones,dual-pulse,electrohydraulic,extracorporeal shock wave lithotripsy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要