[Effects of Low MU in Respiratory Gated IMRT on MLC Position Accuracy and Dose Distribution].

Nihon Hoshasen Gijutsu Gakkai zasshi(2018)

引用 1|浏览0
暂无评分
摘要
It was shown that MLC position accuracy due to stop and go of MLC leaf can be secured even with low MU irradiation of about 3 MU. However, attention should be paid to the dose of risk organs adjacent to the tumor margin.
更多
查看译文
关键词
intensity modulated radiotherapy (IMRT),log-file,low monitor unit (MU),multileaf collimator (MLC),respiratory gated