Performance Characteristics of Commercial Immunoassays for the Detection of IgG and IgM Antibodies to β2 Glycoprotein I and an Initial Assessment of Newly Developed Reference Materials for Assay Calibration.

American journal of clinical pathology(2016)

引用 11|浏览18
暂无评分
摘要
OBJECTIVES:To investigate the performance characteristics and impact of newly developed reference calibrators on the commutability between anti-β2 glycoprotein I (anti-β2 GPI) immunoassays in antiphospholipid syndrome (APS) and/or systemic lupus erythematosus (SLE). METHODS:Immunoglobulin G (IgG) and immunoglobulin M (IgM) anti-β2 GPI immunoassays from four manufacturers were evaluated. Serum samples from 269 patients (APS only, n = 31; SLE and APS, n = 83; SLE only, n = 129; pregnancy-related clinical manifestations without APS, n = 26) and 162 women with histories of successful pregnancies were tested. Results were expressed in kit-specific arbitrary units and in the calibrator reference units (RUs) based on 99th percentile cutoff values. Diagnostic accuracies, correlation between kits, and specific clinical manifestations in APS were investigated. RESULTS:The sensitivities of the assays ranged from 15.8% to 27.2% (IgG) and 12.3% to 15.8% (IgM) while specificities ranged from 79.4% to 86.5% (IgG) and 80.6% to 84.5% (IgM). There was moderate to almost perfect interassay reliability (Cohen κ, 0.69-0.98), and Spearman correlation coefficients were generally improved when results of the IgG determinations were expressed in RUs. CONCLUSIONS:Although qualitative agreements between immunoassays for both antibody isotypes are acceptable, correlations with APS clinical manifestations were kit dependent. Only the use of IgG reference material improved quantitative correlations between assays.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要