Factors Associated With Trunnionosis in the Metal-on-Metal Pinnacle Hip.

The Journal of Arthroplasty(2017)

引用 41|浏览9
暂无评分
摘要
Background: Trunnionosis of the tapered head-stem junction of total hip arthroplasties, either through corrosion or mechanical wear, has been implicated in early implant failure. Retrieval analysis of large numbers of failed implants can help us better understand the factors that influence damage at this interface. Methods: In this study, we examined 120 retrieved total hip arthroplasties of one bearing design, the 36-mm diameter metal-on-metal, DePuy Pinnacle, that had been paired with 3 different stems. We measured material loss of the bearing and head-trunnion taper surfaces and collected clinical and component data for each case. We then used multiple linear regression analysis to determine which factors influenced the rate of taper material loss. Results: We found 4 significant variables: (1) longer time to revision (P = .004), (2) the use of a 12/14 taper for the head-trunnion junction (P < .001), (3) decreased bearing surface wear (P = .003), and (4) vertical femoral offset (P = .05). These together explained 29% of the variability in taper material loss. Conclusion: Our most important finding is the effect of trunnion design. Of the 3 types studied, we found that S-ROM design was the most successful at minimizing trunnionosis. (C) 2016 Elsevier Inc. All rights reserved.
更多
查看译文
关键词
taper,material loss,corrosion,retrieval,metal-on-metal
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要