Digital Breast Tomosynthesis: Physics, Artifacts, and Quality Control Considerations.

RADIOGRAPHICS(2019)

引用 35|浏览9
暂无评分
摘要
As digital breast tomosynthesis (DBT) becomes widely used, radiologists must understand the basic principles of (a) image acquisition, (b) artifacts, and (c) quality control (QC) that are specific to DBT. Standard acquisition parameters common to both full-field digital mammography (FFDM) and DBT are combinations of x-ray tube voltage, current exposure time, and anode target and filter combinations. Image acquisition parameters specific to DBT include tube motion, sweep angle, and number of projections. Continuous tube motion or x-ray emission decreases imaging time but leads to focal spot blurring when compared with step-and-shoot techniques. The sweep angle and number of projections determines resolution. Wider sweep angles allow greater out-of-plane (z-axis) resolution, improving visualization of masses and architecture distortion. A greater number of projections increases in-plane or x-y axis resolution, improving visualization of microcalcifications. Artifacts related to DBT include blurring-ripple, truncation, and loss of skin and superficial tissue resolution. Motion artifacts are difficult to recognize because of inherent out-of-plane blurring. To maintain optimal image quality and an "as low as reasonably achievable" (ALARA) radiation dose, regular QC must be performed. DBT is considered a new imaging modality; therefore, breast imaging facilities are required to obtain a separate certification in addition to that in FFDM, and all personnel (radiologists, technologists, and medical physicists) are mandated to complete initial DBT training and maintain appropriate continuing medical education credits.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要