Preliminary energy-filtering neutron imaging with time-of-flight method on PKUNIFTY: A compact accelerator based neutron imaging facility at Peking University

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2016)

引用 4|浏览12
暂无评分
摘要
Peking University Neutron Imaging Facility (PKUNIFTY) works on an accelerator–based neutron source with a repetition period of 10ms and pulse duration of 0.4ms, which has a rather low Cd ratio. To improve the effective Cd ratio and thus improve the detection capability of the facility, energy-filtering neutron imaging was realized with the intensified CCD camera and time-of-flight (TOF) method. Time structure of the pulsed neutron source was firstly simulated with Geant4, and the simulation result was evaluated with experiment. Both simulation and experiment results indicated that fast neutrons and epithermal neutrons were concentrated in the first 0.8ms of each pulse period; meanwhile in the period of 0.8–2.0ms only thermal neutrons existed. Based on this result, neutron images with and without energy filtering were acquired respectively, and it showed that detection capability of PKUNIFTY was improved with setting the exposure interval as 0.8–2.0ms, especially for materials with strong moderating capability.
更多
查看译文
关键词
Accelerator,Energy-filtering,Neutron imaging,Time-of-flight method