谷歌浏览器插件
订阅小程序
在清言上使用

Evaluation of the Abaxis Piccolo Point-of-Care Chemistry Analyzer: Comparison to the Roche Cobas C501

Point of Care: The Journal of Near-patient Testing & Technology(2017)

引用 1|浏览8
暂无评分
摘要
In this study, we performed method validations for the Abaxis Piccolo Xpress assays for total protein, albumin, alanine aminotransferase, aspartate aminotransferase, alkaline phosphatase, total bilirubin, calcium, creatinine, urea nitrogen, and glucose compared with the Roche Cobas c501 analyzer. The validation included linear reportable range, imprecision, and assessment of accuracy and bias by method crossover. Although some biases between the methods were observed, these could be reconciled by adjustments to the normal reference ranges. Overall, we found that these assays performed at an accepted level for clinical use.
更多
查看译文
关键词
chemistry,point-of-care
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要