Integration Of A Dual-Mode Catheter For Ultrasound Image Guidance And Hifu Ablation Using A 2-D Cmut Array
2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS)(2017)
摘要
Image-guided high-intensity focused ultrasound (HIFU) is widely used not only for non-invasive therapy but also for a precise approach for tissue ablation. Most HIFU systems use piezoelectric transducers, which are typically bulky due to active cooling, and separate imaging and HIFU transducers, and are therefore impractical for catheter-based applications. Taking advantage of a single 2-D capacitive micromachined ultrasonic transducer (CMUT) array, we developed a dual-mode catheter that can switch between ultrasound imaging mode and HIFU ablation mode. The catheter is equipped with an application-specific integrated circuit (ASIC) and a 32 x 32-element 2-D CMUT array. Both ASIC and CMUT are flip-chip bonded to a custom-designed flexible printed circuit board (flex PCB) via 100-mu m and 80-mu m solder balls. Then, the flex legs are folded and terminated with pads for a micro zero insertion force (mu ZIF) connector, allowing easy assembly replacement without the extra cost of coaxial cable assembly. Next, the micro-coaxial cables are assembled at the end of the mu ZIF connectors. After integration with a 3-D printed tip and encapsulating with polydimethylsiloxane (PDMS), the catheter is finalized in a 22-mm diameter shaft. We successfully validated the functionality of both modes of the dual-mode catheter in oil. We are currently preparing the test for an animal study.
更多查看译文
关键词
Dual-Mode Catheter, Integration, flip-chip bonding, Flexible PCB, Ultrasound Image-guided HIFU
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络