Sensitivity Of Cinefluoroscopy Compared With Visual Inspection Of The Extracted Lead For Detection Of Conductor Externalization Of Riata/Riata St Icd Leads

Circulation(2015)

引用 0|浏览7
暂无评分
摘要
Background: The Riata/Riata ST implantable cardioverter defibrillator (ICD) lead series (St Jude Medical, Sylmar, CA) is vulnerable to a characteristic insulation failure mechanism which results in conductor externalization (CE). This may be visible on cinefluoroscopy, however the sensitivity of this test is unknown when compared with direct visible inspection of extracted leads. Methods: We retrospectively reviewed the records of patients with Riata leads extracted at our center since 01/01/2011, and compared CE on fluoroscopic imaging before extraction with the visual appearance of the extracted lead ex vivo. Cinefluoroscopy of the Riata lead was performed in three planes: posteroanterior, left and right anterior oblique. CE was defined as a conductor that had breached the outer insulation material and was clearly visible outside the lead body on fluoroscopy. Ex vivo, CE was defined as externalization of the conductor cable(s) visible on inspection of the extracted lead with a corresponding linear insul...
更多
查看译文
关键词
Riata,ICD lead,Lead failure,Conductor Externalization,cinefluoroscopy