Time resolution measurements with SiPMs coupled to a scintillator

A. Alici, P. Antonioli, S. Arcelli, M. Basile,F. Bellini,F. Carnesecchi, D. Cavazza, L. Cifarelli,F. Cindolo,M. Colocci,A. De Caro,D. De Gruttola,S. De Pasquale,M. Fusco Girard,D. Hatzifotiadou,N. Jacazio,Zhi-Hai Liu,O. Maragoto Rodriguez, A. Margotti,E. Meninno, R. Nania, F. Noferini,O. Pinazza,R. Preghenella,E. Scapparone,G. Scioli,J. Wilkinson, M. C. S. Williams, C. Zampolli, A. Zichichi, R. Zuyeuski

JOURNAL OF INSTRUMENTATION(2018)

引用 3|浏览55
暂无评分
摘要
This paper reports on the results of time resolution measurements of detectors consisting of SiPMs coupled to a scintillator. The R&D has been performed both in a cosmic-ray setup, at the Bologna INFN laboratories, and in a beam test, at the CERN T10 beam line. Different couplings, direct or by means of optical fibres, have been tested. The measurements indicate that to reach better time resolutions, it is important to have a direct coupling between the SiPM and the scintillator. A time resolution of 67 ps has been achieved, for the direct coupling, broadened by the full electronics chain jitter.
更多
查看译文
关键词
Instrumentation and methods for time-of-flight (TOF) spectroscopy,Photon detectors for UV,visible and IR photons (solid-state) (PIN diodes, APDs, Si-PMTS, G-APDs, CCDs, EBCCDs, EMCCDs etc),Scintillators,Scintillators, scintillation and light emission processes (solid, gas and liquid scintillators),Timing detectors
AI 理解论文
溯源树
样例