BACKGROUND AND PURPOSE:Motion management in the treatment of lung cancer is necessary to assure highest quality of the delivered radiation therapy. In this study, the breath-hold technique is experimentally investigated for pencil beam scanned (PBS) proton therapy, with respect to the dosimetric effect of residual breath-hold motion. MATERIAL AND METHODS:Three-dimensional (3D)-printed tumours extracted from CT scans of three patients were inserted into a dynamic anthropomorphic breathing phantom. The target was set up to move with the individual patient's tumour motion during breath-hold as previously assessed on fluoroscopy. Target dose was measured with radio-chromic film, and both single field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT) plans were delivered. Experiments were repeated for each patient without any motion, to compute the relative dose deviation between static and breath-hold cases. RESULTS:SFUD plans showed small dose deviations between static and breath-hold cases, as evidenced by the gamma pass rate (3%, 3 mm) of 85% or higher. Dose deviation was more evident for IMPT plans, with gamma pass rate reduced to 50-70%. CONCLUSIONS:The breath-hold technique is robust to residual intra-breath-hold motion for SFUD treatment plans, based on our experimental study. IMPT was less robust with larger detected dose deviations.