Mechanical Analysis of an MgB2 1.5 T MRI Main Magnet Protected Using Coupling Loss Induced Quench

Cryogenics(2019)

引用 4|浏览16
暂无评分
摘要
Mechanical analysis of the stress and strains developed in the coils were calculated for a ten coil 1.5 T MRI magnet design with magnesium diboride (MgB2) wire protected with Coupling Loss Induced Quench (CLIQ). The temperature distribution inside the coils was first simulated in MATLAB to solve the governing heat and circuit equations. Simulations were performed on the magnet, in which each coil was divided into two subsections, with two CLIQ units while the capacitor ranged from 5 to 20 mF and the initial charging voltage ranged from 2.6 kV to 1.3 kV in order to keep the total stored energy in the CLIQ system constant. The wire's filamentary twist pitch remained constant at 5 cm for all simulations. The exported temperature distribution was expanded to form a representative unit cell (RUC) representing the wire composite and then imported into ANSYS to calculate the 1st principle strain in the MgB2 filament and shear stress across the epoxy for the coils. A peak temperature of 191 K occurred inside the coil with the initial quench when the CLIQ unit had a 20 mF capacitor charged to 1.3 kV. According to the mechanical simulations, the largest resulting peak strain in the wire was 0.034%, and peak shear stress was 44 MPa.
更多
查看译文
关键词
MRI magnet,High temperature superconductor,Numeric simulation,Quench protection,MgB2,Mechanical stress and strain,Coupling loss induced quench