First Results from an Event Synchronized -- High Repetition Thomson Scattering System at Wendelstein 7-X

JOURNAL OF INSTRUMENTATION(2019)

引用 3|浏览101
暂无评分
摘要
The Wendelstein 7-X Thomson scattering diagnostic was upgraded to transiently achieve kilohertz sampling rates combined with adjustable measuring times. The existing Nd:YAG lasers are employed to repetitively emit "bursts", i.e. multiple laser pulses in a short time interval. Appropriately timing bursts in the three available lasers, up to twelve evenly spaced consecutive measurements per burst are possible. The pulse-to-pulse increment within a burst can be tuned from 2 mu s to 33.(3) over bar ms (500 kHz-30 Hz). Additionally, an event trigger system was developed to synchronize the burst Thomson scattering measurements to plasma events. Exemplary, a case of fast electron density and temperature evolution after cryogenic H-2 pellet injection is presented in order to demonstrate the capabilities of the method.
更多
查看译文
关键词
Plasma diagnostics - interferometry, spectroscopy and imaging,Trigger concepts and systems (hardware and software),Lasers,Detector alignment and calibration methods (lasers, sources, particle-beams)