Benchmarking of monte carlo model of siemens oncor® linear accelerator for 18MV photon beam: determination of initial electron beam parameters.

JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY(2020)

引用 3|浏览7
暂无评分
摘要
OBJECTIVE: This study aims to benchmark a Monte Carlo (MC) model of the 18MV photon beam produced by the Siemens Oncor (R) linac using the BEAMnrc and DOSXYZnrc codes. METHODS: By matching the percentage depth doses and beam profiles calculated by MC simulations with measurements, the initial electron beam parameters including electron energy, full width at half maximum (spatial FWHM), and mean angular spread were derived for the 10 x 10 cm(2) and 20 x 20 cm(2) field sizes. The MC model of the 18MV photon beam was then validated against the measurements for different field sizes (5 x 5, 30 x 30 and 40 x 40 cm(2)) by gamma index analysis. RESULTS: The optimum values for electron energy, spatial FWHM and mean angular spread were 14.2MeV, 0.08cm and 0.8 degree, respectively. The MC simulations yielded the comparable measurement results of these optimum parameters. The gamma passing rates (with acceptance criteria of 1%/1 mm) for percentage depth doses were found to be 100% for all field sizes. For cross-line profiles, the gamma passing rates were 100%, 97%, 95%, 96% and 95% for 5 x 5, 10 x 10, 20 x 20, 30 x 30 and 40 x 40 cm(2) field sizes, respectively. CONCLUSIONS: By validation of the MC model of Siemens Oncor (R) linac using various field sizes, it was found that both dose profiles of small and large field sizes were very sensitive to the changes in spatial FWHM and mean angular spread of the primary electron beam from the bending magnet. Hence, it is recommended that both small and large field sizes of the 18MV photon beams should be considered in the Monte Carlo linac modeling.
更多
查看译文
关键词
Radiotherapy,Monte Carlo simulation,benchmarking,Siemens Oncor,gamma analysis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要