An international multi-center serum protein electrophoresis accuracy and M-protein isotyping study. Part I: factors impacting limit of quantitation of serum protein electrophoresis.

CLINICAL CHEMISTRY AND LABORATORY MEDICINE(2020)

引用 32|浏览53
暂无评分
摘要
Background: Serum protein electrophoresis (SPEP) is used to quantify the serum monoclonal component or M-protein, for diagnosis and monitoring of monoclonal gammopathies. Significant imprecision and inaccuracy pose challenges in reporting small M-proteins. Using therapeutic monoclonal antibody-spiked sera and a pooled betamigrating M-protein, we aimed to assess SPEP limitations and variability across 16 laboratories in three continents. Methods: Sera with normal, hypo- or hypergammaglobulinemia were spiked with daratumumab, Dara (cathodal migrating), or elotuzumab, Elo (central-gamma migrating), with concentrations from 0.125 to 10 g/L (n=62) along with a beta-migrating sample (n=9). Provided with total protein (reverse biuret, Siemens), laboratories blindly analyzed samples according to their SPEP and immunofixation (IFE) or immunosubtraction degrees SUM standard operating procedures. Sixteen laboratories reported the perpendicular drop (PD) method of gating the M-protein, while 10 used tangent skimming (TS). A mean percent recovery range of 80%-120% was set as acceptable. The inter-laboratory %CV was calculated. Results: Gamma globulin background, migration pattern and concentration all affect the precision and accuracy of quantifying M-proteins by SPEP. As the background increases, imprecision increases and accuracy decreases leading to overestimation of M-protein quantitation especially evident in hypergamma samples, and more prominent with PD. Cathodal migrating M-proteins were associated with less imprecision and higher accuracy compared to central-gamma migrating M-proteins, which is attributed to the increased gamma background contribution in M-proteins migrating in the middle of the gamma fraction. There is greater imprecision and loss of accuracy at lower M-protein concentrations. Conclusions: This study suggests that quantifying exceedingly low concentrations of M-proteins, although possible, may not yield adequate accuracy and precision between laboratories.
更多
查看译文
关键词
accuracy,immunofixation,immunosubtraction,limit of quantitation,monoclonal proteins,protein electrophoresis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要