Calculated beam quality correction factors for ionization chambers in MV photon beams.

PHYSICS IN MEDICINE AND BIOLOGY(2020)

引用 13|浏览14
暂无评分
摘要
The beam quality correction factor, , which corrects for the difference in the ionization chamber response between the reference and clinical beam quality, is an integral part of radiation therapy dosimetry. The uncertainty of is one of the most significant sources of uncertainty in the dose determination. To improve the accuracy of available data, four partners calculated factors for 10 ionization chamber models in linear accelerator beams with accelerator voltages ranging from 6 MV to 25 MV, including flattening-filter-free (FFF) beams. The software used in the calculations were EGSnrc and PENELOPE, and the ICRU report 90 cross section data for water and graphite were included in the simulations. Volume averaging correction factors were calculated to correct for the dose averaging in the chamber cavities. A comparison calculation between partners showed a good agreement, as did comparison with literature. The values from TRS-398 were higher than our values for each chamber where data was available. The values for the FFF beams did not follow the same , relation as beams with flattening filter (values for 10 MV FFF beams were below fits made to other data on average by 0.3%), although our FFF sources were only for Varian linacs.
更多
查看译文
关键词
beam quality correction factor,K-Q,volume averaging,Monte Carlo,ICRU-90