谷歌浏览器插件
订阅小程序
在清言上使用

The Effect of Roller Head Pump on Platelet Deterioration During the Simulated Extracorporeal Circulation

University of Tsukuba

Journal of artificial organs(2020)

引用 1|浏览24
暂无评分
摘要
Roller pumping results in hemolysis and adverse effects on coagulation, but there are few reports on the influence of roller heads on platelets. Here, we evaluate the interaction between roller pumping and platelet function using a simulated extracorporeal circuit incorporating a vinyl chloride tube and roller head pump with 30 min recirculation. Platelet aggregation, platelet count, microparticle, P-selectin, Phosphatidylserine (PS) exposure and Ricinus Communis Agglutinin 1 (RCA-1) were measured before, 5, 10, 20, and 30 min after the recirculation using 100 ml of fresh human blood that had obtained from healthy volunteers (n = 9). Platelet aggregation and platelet count gradually decreased but microparticles significantly increased after the recirculation (P < 0.05). P-selectin, PS exposure and RCA-1 were measured using flow cytometry. There were no significant differences in the P-selectin and PS exposure expression during recirculation. RCA-1, a platelet apoptosis markers, significantly increased 30 min after recirculation (P < 0.05). We thus conclude that roller pumping induced platelet apoptosis and caused decreases in platelet count and aggregation after the recirculation.
更多
查看译文
关键词
Simulated extracorporeal circuit,Microparticle,P-selectin,PS exposure,RCA-1
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要