3D Myocardial Mechanical Wave Measurements: Toward In Vivo 3D Myocardial Elasticity Mapping

JACC: Cardiovascular Imaging(2021)

引用 19|浏览36
暂无评分
摘要
OBJECTIVES This study aimed to investigate the potential of a novel 3-dimensional (3D) mechanical wave velocity mapping technique, based on the natural mechanical waves produced by the heart itself, to approach a noninvasive 3D stiffness mapping of the left ventricle.BACKGROUND Myocardial fibrosis is recognized as a pathophysiological substrate of major cardiovascular disorders such as cardiomyopathies and valvular heart disease. As fibrosis leads to increased myocardial stiffness, ultrasound elastography measurements could provide important clinical information.METHODS A 3D high frame rate imaging sequence was implemented on a high-end clinical ultrasound scanner to achieve 820 volumes/s when gating over 4 consecutive cardiac cycles. Five healthy volunteers and 10 patients with various degrees of aortic stenosis were included to evaluate feasibility and reproducibility. Mechanical waves were detected using the novel Clutter Filter Wave Imaging approach, shown to be highly sensitive to the weak tissue displacements caused by natural mechanical waves.RESULTS 3D spatiotemporal maps of mechanical wave velocities were produced for all subjects. Only the specific mechanical wave at atrial contraction provided a full 3D coverage of the left ventricle (LV). The average atrial kick propagation velocity was 1.6 +/- 0.2 m/s in healthy volunteers and 2.8 +/- 0.8 m/s in patients (p = 0.0016). A high correlation was found between mechanical wave velocity and age (R-2 = 0.88, healthy group), septal wall thickness (R-2 = 0.73, entire group), and peak jet velocity across the aortic valve (R-2 = 0.70). For 3 of the patients, the higher mechanical wave velocity coexisted with the presence of late gadolinium enhancement on cardiac magnetic resonance.CONCLUSIONS In this study, 3D LV mechanical wave velocities were visualized and measured in healthy volunteers and patients with aortic stenosis. The proposed imaging sequence and measurement technique allowed, for the first time, the measurement of full spatiotemporal 3D elasticity maps of the LV using ultrasound. (C) 2021 The Authors. Published by Elsevier on behalf of the American College of Cardiology Foundation.
更多
查看译文
关键词
left ventricle,natural mechanical wave,stiffness mapping,3D ultrasound high frame rate imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要