Effects of patient orientations, landmark positions, and device positions on the MRI RF-induced heating for modular external fixation devices

MAGNETIC RESONANCE IN MEDICINE(2021)

引用 5|浏览8
暂无评分
摘要
Purpose This paper studies the RF-induced heating for modular external fixation devices applied on the leg regions of the human bodies. Through numerical investigations of RF-induced heating related to different patient orientations, landmark positions, and device positions under 1.5T and 3T MRI systems, simple and practical methods to reduce RF-induced heating are recommended. Methods Numerical simulations using a full-wave electromagnetic solver based on the finite-difference time-domain method were performed to characterize the effects of patient orientations (head-first/feet-first), landmark positions (the scanning area of the patient), and device positions (device on left or right leg) on the RF-induced heating of the external fixation devices. The G32 coil design and three anatomical human models (Duke model, Ella model, and Fats model) were adopted to model the MRI RF coil and the patients. Results The relative positions of the patient, device, and coil can significantly affect the RF-induced heating. With other conditions remaining the same, changing the device position or patient orientation can lead to a peak 1-g averaged spatial absorption ratio variation of a factor around four. By changing the landmark position and the patient orientation, the RF-induced heating can be reduced from 1323.6 W/kg to 217.5 W/kg for the specific scanning situations studied. Conclusion Patient orientations, landmark positions, and device positions influence the RF-induced heating of modular external fixation devices at 1.5 T and 3 T. These features can be used to reduce the RF-induced heating during MRI simply and practically.
更多
查看译文
关键词
device position, landmark position, magnetic resonance imaging, modular external fixation devices, patient orientation, RF-induced heating
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要