Introduction: Turns-amplitude, number of small segments (NSS)-activity, and envelope-activity clouds are three methods of electromyography (EMG) interference pattern analysis. Our objective was to evaluate the sensitivity and specificity of each individual cloud analysis and combined clouds analysis to compare with that of quantitative motor unit potential (QMUP) analysis.Methods: A total of 379 muscles from 100 patients were analyzed by both QMUP and clouds analyses. Calculation of sensitivity and specificity was based on the clinical diagnosis as the "gold standard."Results: For discrimination of abnormal vs normal and neuropathic vs non-neuropathic, combined clouds analysis had greater sensitivity than QMUP analysis and any single cloud analysis, but there were no differences in specificity. For discrimination of myopathic vs non-myopathic, combined clouds analysis and single cloud analysis had greater sensitivity than QMUP analysis, but there were no differences in specificity.Discussion: Combined clouds analysis was superior to QMUP and each single cloud analysis for distinguishing normal, myopathic, and neuropathic muscles.
更多
查看译文
关键词
electromyography, envelope-activity cloud, interference pattern analysis, NSS-activity cloud, quantitative motor unit potential, turns-amplitude cloud