64-channel photon-counting computed tomography using a new MPPC-CT system

H. Kiji, T. Maruhashi, T. Toyoda,J. Kataoka,M. Arimoto,D. Sato, K. Yoshiura,S. Kobayashi, H. Kawashima, S. Terazawa, S. Shiota,H. Ikeda

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2020)

引用 10|浏览27
暂无评分
摘要
X-ray computed tomography (CT) is a widely used diagnostic tool to visualize the interior of the human body. However, the exposure dose of conventional CT in a single scan is large, typically 10 mSv, and therefore, it is necessary to find ways to reduce the radiation dose. Furthermore, conventional CT does not contain the energy information of individual X-ray photons because the X-ray signals are read out as an integrated form. This causes misidentification of materials. To resolve this issue, we propose a novel photon counting CT (PC-CT) system consisting of multi-pixel photon counters (MPPCs) coupled with high speed scintillators. The system has a 64-channel MPPC array that improves energy information and wide-area imaging. By fine energy adjustment and increasing the number of energy thresholds to six, which were newly implemented in the 64-channel PC-CT system, we succeeded in accurately estimating the concentrations of contrast agents such as iodine and gadolinium. Moreover, for mixed phantoms of iodine and gadolinium, we demonstrate discrimination between them, and estimate the concentrations individually, which cannot be done by conventional CTs. This shows great potential in expanding the applications of X-ray CTs.
更多
查看译文
关键词
Photon counting computed tomography,Multipixel photon counter,K-edge imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要