Evaluation Of High-Linearity Bone Radiation Detectors Exposed To Gamma-Rays Via Ftir Measurements

APPLIED RADIATION AND ISOTOPES(2021)

引用 0|浏览1
暂无评分
摘要
In radiation physics, the study of new alternative dosimeters is of interest to the growing branch of dosimetric characterization for radiotherapy applications. The goal of this work was to expose bone samples to high doses and evaluate their linearity response to gamma rays. The Fourier Transform Infrared (FTIR) spectrophotometry technique was employed as the evaluation technique, and based on the spectrophotometry absorbance profiles the linearity was assessed based on the following methods: Area Under the Curve (AUC), Wavenumber Method (WM), Partial Component Regression (PCR) and Partial Least-Square Regression (PLSR) methods. The bone samples were irradiated with absorbed doses of 10 Gy up to 500 Gy using a Co-60 Gamma Cell-220 system. The results showed, for the calibration curves of the system, adequate linearity on all methods. In conclusion, the results indicate a good linear response and therefore an interesting potential radiation detector.
更多
查看译文
关键词
Bone, Radiation dosimetry, PLSR and PCR analyses, FTIR technique, High doses