SARS-CoV-2 (Covid-19) workplace temperature screening: Seasonal concerns for thermal detection in northern regions.

Applied ergonomics(2021)

引用 13|浏览12
暂无评分
摘要
Workplace temperature screening has become standard practice during the SARS-CoV-2 pandemic. The objective was to determine the consistency of four temperature devices during exposure to simulated and actual environmental conditions reflective of a workplace. An infrared (IR) digital thermometer (accuracy(A)±0.2), IR laser thermometer (A±1), and thermal imaging camera (A±0.3) were used to measure forehead and tympanic (digital only) temperatures. The first experiment was conducted in a controlled simulated environment (-20 to 20 °C) with three participants (32-YOF, 27-YOM, 20-YOF). The second experiment used actual outdoor conditions (-0.48 to 45.6 °C) with two participants (32-YOF, 27-YOM). The tympanic measurement was the least impacted by environmental temperature (mean(±SD)): simulated (36.8(±0.18) °C) and actual (36.9(±0.16) °C). The thermal imaging camera had the lowest RMSE values (0.81-0.97 °C), with outdoor temperatures ranging from 0 to 45 °C. Environmental temperature influenced forehead temperature readings and required a resting period in a thermoneutral environment (5-9 min (-20 to -10 °C) to immediate (15-20 °C)).