We developed the image processing system using SISR based on CNN to reduce the CBCT imaging dose in radiation therapy. The developed system could add the high-frequency components which were not included in low matrix size CBCT images. CBCT imaging dose could be reduced to 1/16 and 1/4 when CBCT scans were taken with 128×128 and 256×256 pixels, respectively. The developed system enables CBCT imaging using a low radiation dose without degrading the image quality.