Assessing The Impact Of Nonspecific Binding On Oligonucleotide Bioanalysis

BIOANALYSIS(2021)

引用 12|浏览11
暂无评分
摘要
Aim: Accurate and reliable quantification of oligonucleotides can be difficult, which has led to an increased focus on bioanalytical methods for more robust analyses. Recent advances toward mitigating sample losses on liquid chromatography (LC) systems have produced recovery advantages for oligonucleotide separations. Results & methodology: LC instruments and columns constructed from MP35N metal alloy and stainless steel columns were compared against LC hardware modified with hybrid inorganic-organic silica surfaces. Designed to minimize metal-analyte adsorption, these surfaces demonstrated a 73% increase in 25-mer phosphorothioate oligonucleotide recovery using ion-pairing reversed-phase LC versus standard LC surfaces, most particularly upon initial use. Conclusion: Hybrid silica chromatographic surfaces improve the performance, detection limits and reproducibility of oligonucleotide bioanalytical assays.
更多
查看译文
关键词
bioanalysis, ion-pairing, LC-UV-MS, mass spectrometry, oligonucleotides, pH, phosphorothioate, spectrofluorometric quantitation