Biomechanical Performance of BoneHelix(R) Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model

Laura Leonie Brandes,Luis Fernando Nicolini,Johannes Greven,Philipp Lichte, Thomas Thaddäus Stopinski, Martin Sattler,Frank Hildebrand,Miguel Pishnamaz

LIFE-BASEL(2021)

引用 3|浏览0
暂无评分
摘要
Tibial shaft fractures are common injuries in the pediatric and adolescent populations. Elastic stable intramedullary nailing (ESIN) is the treatment of choice for cases that require surgical stabilization. A new intramedullary device, BoneHelix(R) (BH), may be an alternative for use with fractures that cannot be satisfactorily stabilized with ESIN. This study aimed to assess the biomechanical performance of BH compared with ESIN in a porcine tibia fracture model, observing cyclic fatigue and load to failure. Computed tomography was used to monitor the implant position and to rule out unintended damage. No implant or bone failure occurred during the fatigue testing. An increase in the cumulative plastic displacement was observed in both test groups over the loading cycles applied. Both implant-bone constructs displayed a trend toward closure of the osteotomy gap. During the load-to-failure test, the average loads at failure in specimens instrumented with ESIN and BH were 5364 N (& PLUSMN;723) and 4350 N (& PLUSMN;893), respectively, which were not statistically significant (p = 0.11). The values of both groups were two to three times higher than the estimated maximal load (2000 N) during physiological weight bearing. The biomechanical results thus indicate equivalent performance and stability by the implants tested.
更多
查看译文
关键词
fracture, tibia, pediatric, cyclic testing, elastic stable intramedullary nailing (ESIN), BoneHelix(R), biomechanics
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要