Design of corneal cross-linking system: with real-time positioning of pupil position

IOP Conference Series: Earth and Environmental Science(2019)

引用 0|浏览1
暂无评分
摘要
Abstract This paper presents a design of a corneal cross-linking instrument with the function of tracking and locating the central position of the pupil in real time. According to the application requirements, based on the principle of uniform light of compound eye lens, the optical projection system of the corneal crosslinking instrument with the diameter of the treatment spot being 9mm and the uniformity being greater than 90% was designed. And put forward a kind of the pupil localization algorithm, the acquisition of the pupil image gray processing and calculate the histogram, according to the characters of histogram adaptively to take out the appropriate threshold binarization, again after the binarization of image contour tracking, get accurate boundary, the outline of the final vote based on the sliding arc length increment method to determine the pupil center and radius. Experimental results show that the corneal cross-linking system designed in this paper can not only accurately project the treatment of light spot, but also accurately track the central position of the pupil of the human eye.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper